Help on: How Many Times can System Suitability Tests be Repeated?

MTS HELPDESK
Do you have any problems relating to analytical chemistry for pharmaceuticals or training? Send your questions to the MTS helpdesk using our contact form.
Question:
“How many times can a system suitability test for an HPLC method be repeated?”
Answer:
“If your question refers to repeating the system suitability test after a failure without making any changes to the system then the number of times this can be done is zero. The reasoning for this statement is as follows: The purpose of the test is to check that the method is performing as expected, if it fails then it indicates that there is a problem and then the source of that problem needs to be identified and corrected before the system suitability can be performed again. Admittedly, on many occasions only a probable cause can be identified and an investigation will involve making a change and then using the system suitability test to see it the problem has been corrected. Therefore, provided that there is a justifiable reason (some type of hypothesis about what has caused the failure) each time the test is carried out, I can see no reason why there should be a limit of the number of times the system suitability test is performed. The aim is to ensure that the method is completely controlled by the analyst.”
  